Advanced Benchtop Oximeter – HI6421

REF : HI6421-01

  • Measuring range: 0.00 to 90.00 mg/L (ppm); Saturation 0.0 to 600.0% saturation
  • Resolution: 0.01 mg/L (ppm)
  • Accuracy: 0.00 to 20.00 mg/L (ppm) 1.5% / 20.00 to 50.00 mg/L (ppm) ±5%
  • Temperature range: −20.0 to 120.0 °C
  • Resolution: 0.1°C
  • 7-inch capacitive touchscreen with multi-touch support
  • Recorded data can be exported via Ethernet/Wi-Fi connection and transferred via email and FTP
REQUEST A FREE QUOTE

DESCRIPTION

Compact and easy to use:

HI6421 incorporates Hanna's HI7641133 optical dissolved oxygen (opdo) probe based on the fluorescence quenching principle. An immobilized Pt-based phosphor is excited by light from a blue LED and emits red light. As oxygen interacts with the phosphor, it reduces the intensity and lifetime of the luminescence. The luminescence lifetime is measured by a photodetector and is used to calculate the dissolved oxygen concentration.®
  • 7-inch capacitive touchscreen with multi-touch support
  • Capacitive touch feedback, home and system menu keys
  • User-friendly icons and symbols allow users to easily navigate and interpret the instrument's functions.

Calibration:

One or two calibration points at 0% and/or 100% saturation. Manual one-time calibration in mg/L or % saturation using a reference method. Non-volatile memory stores data and settings

Zero Oxygen Calibration Solution:

When performing a DO calibration, it is important that the correct standards are used. We include a starter solution to help you get started. HI7040L is a two-component, zero-oxygen solution designed so you can quickly calibrate to 0% oxygen concentration.

Connectivity Features and Services:

  • Transfer the saved data to a USB flash drive
  • Log files containing measurements and calibration data (as a .csv file)
  • FTP and email for exporting logs via Ethernet and Wi-Fi connection
  • USB type A for USB flash drive, keyboard and printer / USB type C for USB flash drive and PC connection

Measure :

  • Application-specific profiles enable fast and direct measurement without the need to update sensor and system parameters
  • Active log during measurement
  • Measurement stability indicator (using the Stability Criteria parameter)
  • Playback modes: direct and direct/autohold
  • Temperature compensation can be automatic or manually set
  • Audible and/or alarm messages for measurements outside predefined limits
  • Galvanic isolation for measurement